== 3.0.21 (07-Dec-2009) ==
 
 
 
'''Changes'''
 
*Now allows 0.1s Averaging in FMS mode
 
 
 
'''Bug Fixes'''
 
*PTOF Display bug fixed, first display was not working in v3019
 
*Nitrate Loading value fixed, displayed value had error in v3019
 
*Swap button bug fixed, Calibration Group and Sampling Type values were lost when swap was used in previous versions
 
 
 
== 3.0.19 (29-Sept-2009)==
 
 
 
'''CHANGES'''
 
 
 
'''InfoVals'''
 
*InfoVal(106) indicates seconds after midnight at start of run
 
*InfoVal(107) indicates active menu number
 
*If DIO5, 6, or 7 is in, save state as save
 
*InfoVal(108, 109, 110) states of digital in 5, 6, 7 (NaN if disabled)
 
*InfoVal(111) Relative time within fast MS cycle (NaN if not FMS mode)
 
*InfoVal(112) Relative run number within FMS cycle (NaN if not FMS mode)
 
 
 
'''Menu Switching'''
 
*New menu value allows optional wait at the start of a run
 
*Added Menu Cycle Time Grid as alternative to run-based time grid
 
 
 
'''BitWise'''
 
*Baseline and SI area now based on MS mode; no longer need chopper to run Bitwise
 
*Single Ion peak list is now saved to menu file
 
*Optional "Bitwise on the hour" measures baseline and SI area every hour
 
*All three displayed nominal masses share a single peak discriminator, saved as menuVal
 
*Suggests best baseline bit, see: http://tof-ams.blogspot.com/2009/08/what-is-proper-baseline-bit-value.html
 
 
 
'''Acquisition / Acquisition Window'''
 
*Logic allows FMS with closed block = 0
 
*Logic allows MS mode with % Open = 100
 
*Added functionality and interface updates for MS acquisition without chopper (controlled with menu val 141)
 
*Zoom windows on right side of acquisition window allow change of nominal mass with scroll wheel
 
*If MSclosed does not exist, displayed resolution values are calculated from open data
 
*Option to display MSClosed sticks
 
*Horizontal zoom in stick display with mouse
 
*Eliminated %MSOpen. MS Dwell time now calculated as (MSOpenT+MSClosedT)*MSLoopsPerDwell.  Loops enables multiple MSClosed / MSOpen *Dwells per dwell.
 
*Vertical scaling in zoom windows
 
*Check RealTime Configuration at load of window.  See [[http://cires.colorado.edu/jimenez-group/wiki/index.php/ToF-AMS_DAQ_FAQs#RealTime_Warning_Pop-up | Related FAQ]]
 
*PTOF DC markers calculation adjusted based on number of TOF/chop and co-adds
 
*Fixed display of speciated sticks for SP-AMS Dark Green for Cn and Fullerene
 
*MS Stick display size depends on what other modes are active in acquisition cycle
 
 
 
'''Digital Lines'''
 
*DIO5, 6, and 7 now configurable as in or out (common across menu)
 
If out, state unique to each menu
 
If in, value is recorded and saved as infoVal
 
*Option to control acquisition start/stop from DIO5
 
*Option to control menu switching with encoding on 6 and 7
 
 
 
'''Mass Calibration Window'''
 
*Peak intensity displayed in units of Hz; calculation uses bit-ns per second, SI Area, and TOF pulser frequency
 
*Total Ion Current (TIC) reported below trace of peak areas/heights
 
*Option to save itx file with last acquired mass spectrum.  Includes raw MS, raw mass axis, and unit mass resolution stick spectrum
 
*Added Accumulation mode as option on tuning tab.  MS data are continuously acquired and summed, allowing determination of resolution (etc) for low intensity peaks.
 
 
 
'''Window Messages'''
 
*Allows control of acquire start/stop and menu number from external software with windows messages
 
 
 
'''Real Time Files / Serial Output'''
 
*Write MSSDiff.itx to C:\ToFAMS\ToFAMSData\LogFiles\TimeTrace\MS for each run.  Optional - controlled by check box in Real Time box on averagin and saving tab.
 
*New Serial output formats
 
 
 
'''Front Panel'''
 
*If DAQ PC has internet connection, check for updated version of DAQ.  If update exists, button on panel links to download.
 
 
 
'''National Instruments'''
 
*Auto-determination of National instrument board device numbers.  If board(s) not found, features disabled accordingly.
 
 
 
== 3.0.3 ==
 
'''NEW FEATURES since 3.0.0:'''
 
 
 
'''Acquisition Window'''
 
 
 
Optional automatic baseline and SI determination during acquisition.  Turned on/off on AP240 tab of Menu Window.  This is still considered a BETA feature. Last values are saved as InfoVals.
 
 
 
'''Calib Window
 
'''
 
Step sizes (volts) for TPS scans are saved as common menu values
 
 
 
Scroll wheel and arrow keys to change TPS voltages and calibration center points
 
 
 
'''Info Vals'''
 
If TPS remote control is enabled, last monitored TPS voltages are saved as infoVals for each run
 
 
 
'''Data Save'''
 
 
 
Allows transient save in FMS mode.  itx written to transient drive, without write to C: drive
 
 
 
 
 