== 3.0.24 (07-Jan-2009) ==
 
 
 
'''CHANGES'''
 
*'''Light Scattering'''
 
** DAQ version written as attribute of PartInfo dataset
 
** Second Channel NI6024 Acquisition re-enabled (was disabled with new timing scheme in 3.0.23)
 
 
 
*'''Front Panel'''
 
** Help buttons moved to control menu
 
 
 
'''FIXES'''
 
*'''Bitwise'''
 
** +/- 10V changes in MCP now saved to menu file
 
 
 
== 3.0.23 (29-Dec-2009) ==
 
 
 
'''CHANGES'''
 
*'''Light Scattering'''
 
** New timing scheme for synchronizing PTOF (AP240) and LS (NI6024) data
 
 
 
== 3.0.22 (29-Dec-2009) ==
 
 
 
'''CHANGES'''
 
*'''Light Scattering'''
 
** Add 3 diagnostic values to partInfo: 
 
***(9) Chop Number at Download (Chem minus LS)
 
***(13) Wait for LS Download (s)
 
***(14) T at Download (Chem - LS)
 
 
 
 
 
== 3.0.21 (07-Dec-2009) ==
 
 
 
'''CHANGES'''
 
 
 
*'''Acquisition'''
 
**Now allows 0.1s Averaging in FMS mode
 
 
 
'''FIXES'''
 
*PTOF Display bug fixed, first display was not working in v3019
 
*Nitrate Loading value fixed, displayed value had error in v3019
 
*Swap button bug fixed, Calibration Group and Sampling Type values were lost when swap was used in previous versions
 
 
 
== 3.0.19 (29-Sept-2009)==
 
 
 
'''CHANGES'''
 
 
 
'''InfoVals'''
 
*InfoVal(106) indicates seconds after midnight at start of run
 
*InfoVal(107) indicates active menu number
 
*If DIO5, 6, or 7 is in, save state as save
 
*InfoVal(108, 109, 110) states of digital in 5, 6, 7 (NaN if disabled)
 
*InfoVal(111) Relative time within fast MS cycle (NaN if not FMS mode)
 
*InfoVal(112) Relative run number within FMS cycle (NaN if not FMS mode)
 
 
 
'''Menu Switching'''
 
*New menu value allows optional wait at the start of a run
 
*Added Menu Cycle Time Grid as alternative to run-based time grid
 
 
 
'''BitWise'''
 
*Baseline and SI area now based on MS mode; no longer need chopper to run Bitwise
 
*Single Ion peak list is now saved to menu file
 
*Optional "Bitwise on the hour" measures baseline and SI area every hour
 
*All three displayed nominal masses share a single peak discriminator, saved as menuVal
 
*Suggests best baseline bit, See  [[http://cires.colorado.edu/jimenez-group/wiki/index.php/ToF-AMS_DAQ_FAQs#What_is_the_proper_Baseline_bit_value | related FAQ]]
 
 
 
'''Acquisition / Acquisition Window'''
 
*Logic allows FMS with closed block = 0
 
*Logic allows MS mode with % Open = 100
 
*Added functionality and interface updates for MS acquisition without chopper (controlled with menu val 141)
 
*Zoom windows on right side of acquisition window allow change of nominal mass with scroll wheel
 
*If MSclosed does not exist, displayed resolution values are calculated from open data
 
*Option to display MSClosed sticks
 
*Horizontal zoom in stick display with mouse
 
*Eliminated %MSOpen. MS Dwell time now calculated as (MSOpenT+MSClosedT)*MSLoopsPerDwell.  Loops enables multiple MSClosed / MSOpen *Dwells per dwell.
 
*Vertical scaling in zoom windows
 
*Check RealTime Configuration at load of window.  See [[http://cires.colorado.edu/jimenez-group/wiki/index.php/ToF-AMS_DAQ_FAQs#RealTime_Warning_Pop-up | Related FAQ]]
 
*PTOF DC markers calculation adjusted based on number of TOF/chop and co-adds
 
*Fixed display of speciated sticks for SP-AMS Dark Green for Cn and Fullerene
 
*MS Stick display size depends on what other modes are active in acquisition cycle
 
 
 
'''Digital Lines'''
 
*DIO5, 6, and 7 now configurable as in or out (common across menu)
 
If out, state unique to each menu
 
If in, value is recorded and saved as infoVal
 
*Option to control acquisition start/stop from DIO5
 
*Option to control menu switching with encoding on 6 and 7
 
 
 
'''Mass Calibration Window'''
 
*Peak intensity displayed in units of Hz; calculation uses bit-ns per second, SI Area, and TOF pulser frequency
 
*Total Ion Current (TIC) reported below trace of peak areas/heights
 
*Option to save itx file with last acquired mass spectrum.  Includes raw MS, raw mass axis, and unit mass resolution stick spectrum
 
*Added Accumulation mode as option on tuning tab.  MS data are continuously acquired and summed, allowing determination of resolution (etc) for low intensity peaks.
 
 
 
'''Window Messages'''
 
*Allows control of acquire start/stop and menu number from external software with windows messages
 
 
 
'''Real Time Files / Serial Output'''
 
*Write MSSDiff.itx to C:\ToFAMS\ToFAMSData\LogFiles\TimeTrace\MS for each run.  Optional - controlled by check box in Real Time box on averagin and saving tab.
 
*New Serial output formats
 
 
 
'''Front Panel'''
 
*If DAQ PC has internet connection, check for updated version of DAQ.  If update exists, button on panel links to download.
 
 
 
'''National Instruments'''
 
*Auto-determination of National instrument board device numbers.  If board(s) not found, features disabled accordingly.
 
 
 
== 3.0.3 ==
 
'''CHANGES''' (Since 3.0.0)
 
 
 
'''Acquisition Window'''
 
*Optional automatic baseline and SI determination during acquisition.  Turned on/off on AP240 tab of Menu Window.  This is still considered a BETA feature. Last values are saved as InfoVals.
 
 
 
'''Calib Window '''
 
*Step sizes (volts) for TPS scans are saved as common menu values
 
*Scroll wheel and arrow keys to change TPS voltages and calibration center points
 
 
 
'''Info Vals'''
 
*If TPS remote control is enabled, last monitored TPS voltages are saved as infoVals for each run
 
 
 
'''Data Save'''
 
*Allows transient save in FMS mode.  itx written to transient drive, without write to C: drive
 
 
 