− | == 3.0.21 (07-Dec-2009) ==
| |
− | | |
− | '''Changes'''
| |
− | | |
− | Now allows 0.1s Averaging in FMS mode
| |
− | | |
− | '''Bug Fixes'''
| |
− | | |
− | PTOF Display bug fixed, first display was not working in v3019
| |
− | | |
− | Nitrate Loading value fixed, displayed value had error in v3019
| |
− | | |
− | Swap button bug fixed, Calibration Group and Sampling Type values were lost when swap was used in previous versions
| |
− | | |
− | == 3.0.19 (29-Sept-2009)==
| |
− | | |
− | '''InfoVals'''
| |
− | | |
− | InfoVal(106) indicates seconds after midnight at start of run
| |
− | | |
− | InfoVal(107) indicates active menu number
| |
− | | |
− | If DIO5, 6, or 7 is in, save state as save
| |
− | | |
− | InfoVal(108, 109, 110) states of digital in 5, 6, 7 (NaN if disabled)
| |
− | | |
− | InfoVal(111) Relative time within fast MS cycle (NaN if not FMS mode)
| |
− | | |
− | InfoVal(112) Relative run number within FMS cycle (NaN if not FMS mode)
| |
− | | |
− | '''Menu Switching'''
| |
− | | |
− | New menu value allows optional wait at the start of a run
| |
− | | |
− | Added Menu Cycle Time Grid as alternative to run-based time grid
| |
− | | |
− | '''BitWise'''
| |
− | | |
− | Baseline and SI area now based on MS mode; no longer need chopper to run Bitwise
| |
− | | |
− | Single Ion peak list is now saved to menu file
| |
− | | |
− | Optional "Bitwise on the hour" measures baseline and SI area every hour
| |
− | | |
− | All three displayed nominal masses share a single peak discriminator, saved as menuVal
| |
− | | |
− | Suggests best baseline bit, see: http://tof-ams.blogspot.com/2009/08/what-is-proper-baseline-bit-value.html
| |
− | | |
− | | |
− | '''Acquisition / Acquisition Window'''
| |
− | | |
− | Logic allows FMS with closed block = 0
| |
− | | |
− | Logic allows MS mode with % Open = 100
| |
− | | |
− | Added functionality and interface updates for MS acquisition without chopper (controlled with menu val 141)
| |
− | | |
− | Zoom windows on right side of acquisition window allow change of nominal mass with scroll wheel
| |
− | | |
− | If MSclosed does not exist, displayed resolution values are calculated from open data
| |
− | | |
− | Option to display MSClosed sticks
| |
− | | |
− | Horizontal zoom in stick display with mouse
| |
− | | |
− | Eliminated %MSOpen. MS Dwell time now calculated as (MSOpenT+MSClosedT)*MSLoopsPerDwell. Loops enables multiple MSClosed / MSOpen Dwells per dwell.
| |
− | | |
− | Vertical scaling in zoom windows
| |
− | | |
− | Check RealTime Configuration at load of window. See: http://tof-ams.blogspot.com/2009/08/quickview-warning-window.html
| |
− | | |
− | PTOF DC markers calculation adjusted based on number of TOF/chop and co-adds
| |
− | | |
− | Fixed display of speciated sticks for SP-AMS Dark Green for Cn and Fullerene
| |
− | | |
− | MS Stick display size depends on what other modes are active in acquisition cycle
| |
− | | |
− | '''Digital Lines'''
| |
− | | |
− | DIO5, 6, and 7 now configurable as in or out (common across menu)
| |
− | | |
− | If out, state unique to each menu
| |
− | | |
− | If in, value is recorded and saved as infoVal
| |
− | | |
− | Option to control acquisition start/stop from DIO5
| |
− | | |
− | Option to control menu switching with encoding on 6 and 7
| |
− | | |
− | '''Mass Calibration Window'''
| |
− | | |
− | Peak intensity displayed in units of Hz; calculation uses bit-ns per second, SI Area, and TOF pulser frequency
| |
− | | |
− | Total Ion Current (TIC) reported below trace of peak areas/heights
| |
− | | |
− | Option to save itx file with last acquired mass spectrum. Includes raw MS, raw mass axis, and unit mass resolution stick spectrum
| |
− | | |
− | Added Accumulation mode as option on tuning tab. MS data are continuously acquired and summed, allowing determination of resolution (etc) for low intensity peaks.
| |
− | | |
− | '''Window Messages'''
| |
− | | |
− | Allows control of acquire start/stop and menu number from external software with windows messages
| |
− | | |
− | '''Real Time Files / Serial Output'''
| |
− | | |
− | Write MSSDiff.itx to C:\ToFAMS\ToFAMSData\LogFiles\TimeTrace\MS for each run. Optional - controlled by check box in Real Time box on averagin and saving tab.
| |
− | | |
− | New Serial output formats
| |
− | | |
− | '''Front Panel'''
| |
− | | |
− | If DAQ PC has internet connection, check for updated version of DAQ. If update exists, button on panel links to download.
| |
− | | |
− | '''National Instruments'''
| |
− | | |
− | Auto-determination of National instrument board device numbers. If board(s) not found, features disabled accordingly.
| |
− | | |
− | == 3.0.3 ==
| |
− | '''NEW FEATURES since 3.0.0:'''
| |
− | | |
− | '''Acquisition Window'''
| |
− | | |
− | Optional automatic baseline and SI determination during acquisition. Turned on/off on AP240 tab of Menu Window. This is still considered a BETA feature. Last values are saved as InfoVals.
| |
− | | |
− | '''Calib Window
| |
− | '''
| |
− | Step sizes (volts) for TPS scans are saved as common menu values
| |
− | | |
− | Scroll wheel and arrow keys to change TPS voltages and calibration center points
| |
− | | |
− | '''Info Vals'''
| |
− | If TPS remote control is enabled, last monitored TPS voltages are saved as infoVals for each run
| |
− | | |
− | '''Data Save'''
| |
− | | |
− | Allows transient save in FMS mode. itx written to transient drive, without write to C: drive
| |
− | | |
− | | |