Jimenez Lab Troubleshooting

From Jimenez Group Wiki
Revision as of 11:36, 14 October 2009 by Mlechner2 (talk | contribs)
Jump to: navigation, search

Jimenez Lab Troubleshooting Techniques

Gilibrator

  • Unit will not turn on
    • remove flow cell
    • open back cover
    • unplug battery pack from main board
    • plug in power supply
    • If unit turns on battery pack is faulty, if unit does not turn on board is faulty.
    • Ordering battery pack: Reis environmental 1.800.677.7347 p/n 700560