ToF-AMS DAQ Release Notes
From Jimenez Group Wiki
See Also: ToF-AMS DAQ; ToF-AMS DAQ FAQs; ToF-AMS Main
Contents
3.0.24 (07-Jan-2010)
CHANGES
- Light Scattering
- DAQ version written as attribute of PartInfo dataset
- Second Channel NI6024 Acquisition re-enabled (was disabled with new timing scheme in 3.0.23)
- Front Panel
- Help buttons moved to control menu
FIXES
- Bitwise
- +/- 10V changes in MCP now saved to menu file
3.0.23 (29-Dec-2009)
CHANGES
- Light Scattering
- New timing scheme for synchronizing PTOF (AP240) and LS (NI6024) data
3.0.22 (29-Dec-2009)
CHANGES
- Light Scattering
- Add 3 diagnostic values to partInfo:
- (9) Chop Number at Download (Chem minus LS)
- (13) Wait for LS Download (s)
- (14) T at Download (Chem - LS)
- Add 3 diagnostic values to partInfo:
3.0.21 (07-Dec-2009)
CHANGES
- Acquisition
- Now allows 0.1s Averaging in FMS mode
FIXES
- PTOF Display bug fixed, first display was not working in v3019
- Nitrate Loading value fixed, displayed value had error in v3019
- Swap button bug fixed, Calibration Group and Sampling Type values were lost when swap was used in previous versions
3.0.19 (29-Sept-2009)
CHANGES
InfoVals
- InfoVal(106) indicates seconds after midnight at start of run
- InfoVal(107) indicates active menu number
- If DIO5, 6, or 7 is in, save state as save
- InfoVal(108, 109, 110) states of digital in 5, 6, 7 (NaN if disabled)
- InfoVal(111) Relative time within fast MS cycle (NaN if not FMS mode)
- InfoVal(112) Relative run number within FMS cycle (NaN if not FMS mode)
Menu Switching
- New menu value allows optional wait at the start of a run
- Added Menu Cycle Time Grid as alternative to run-based time grid
BitWise
- Baseline and SI area now based on MS mode; no longer need chopper to run Bitwise
- Single Ion peak list is now saved to menu file
- Optional "Bitwise on the hour" measures baseline and SI area every hour
- All three displayed nominal masses share a single peak discriminator, saved as menuVal
- Suggests best baseline bit, See [| related FAQ]
Acquisition / Acquisition Window
- Logic allows FMS with closed block = 0
- Logic allows MS mode with % Open = 100
- Added functionality and interface updates for MS acquisition without chopper (controlled with menu val 141)
- Zoom windows on right side of acquisition window allow change of nominal mass with scroll wheel
- If MSclosed does not exist, displayed resolution values are calculated from open data
- Option to display MSClosed sticks
- Horizontal zoom in stick display with mouse
- Eliminated %MSOpen. MS Dwell time now calculated as (MSOpenT+MSClosedT)*MSLoopsPerDwell. Loops enables multiple MSClosed / MSOpen *Dwells per dwell.
- Vertical scaling in zoom windows
- Check RealTime Configuration at load of window. See [| Related FAQ]
- PTOF DC markers calculation adjusted based on number of TOF/chop and co-adds
- Fixed display of speciated sticks for SP-AMS Dark Green for Cn and Fullerene
- MS Stick display size depends on what other modes are active in acquisition cycle
Digital Lines
- DIO5, 6, and 7 now configurable as in or out (common across menu)
If out, state unique to each menu If in, value is recorded and saved as infoVal
- Option to control acquisition start/stop from DIO5
- Option to control menu switching with encoding on 6 and 7
Mass Calibration Window
- Peak intensity displayed in units of Hz; calculation uses bit-ns per second, SI Area, and TOF pulser frequency
- Total Ion Current (TIC) reported below trace of peak areas/heights
- Option to save itx file with last acquired mass spectrum. Includes raw MS, raw mass axis, and unit mass resolution stick spectrum
- Added Accumulation mode as option on tuning tab. MS data are continuously acquired and summed, allowing determination of resolution (etc) for low intensity peaks.
Window Messages
- Allows control of acquire start/stop and menu number from external software with windows messages
Real Time Files / Serial Output
- Write MSSDiff.itx to C:\ToFAMS\ToFAMSData\LogFiles\TimeTrace\MS for each run. Optional - controlled by check box in Real Time box on averagin and saving tab.
- New Serial output formats
Front Panel
- If DAQ PC has internet connection, check for updated version of DAQ. If update exists, button on panel links to download.
National Instruments
- Auto-determination of National instrument board device numbers. If board(s) not found, features disabled accordingly.
3.0.3
CHANGES (Since 3.0.0)
Acquisition Window
- Optional automatic baseline and SI determination during acquisition. Turned on/off on AP240 tab of Menu Window. This is still considered a BETA feature. Last values are saved as InfoVals.
Calib Window
- Step sizes (volts) for TPS scans are saved as common menu values
- Scroll wheel and arrow keys to change TPS voltages and calibration center points
Info Vals
- If TPS remote control is enabled, last monitored TPS voltages are saved as infoVals for each run
Data Save
- Allows transient save in FMS mode. itx written to transient drive, without write to C: drive
Archive
Archive of release notes from v2.1.4